Remote Control Commands
R&S
®
SMF100A
552
Operating Manual 1167.2319.02 ─ 12
Example:
ROSC:EFC:STAT ON
Activates the electronic frequency control.
Manual operation:
See
[:SOURce]:ROSCillator:EXTernal:FREQuency
<Frequency>
Informs the instrument of the frequency of the external reference. Entries are possible
in 1 MHz steps.
Parameters:
<Frequency>
float
Range:
1E6 to 2E7
Increment: 1E6
*RST:
1E7
Example:
ROSC:SOUR EXT
Selects the external source. The reference must be input at the
REF IN input connector.
ROSC:EXT:FREQ 5 MHz
Informs the instrument that the external reference frequency is 5
MHz.
Manual operation:
See
"External Reference Frequency"
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
<State>
Activates that RF output is automatically switched off, when in external source mode
no reference signal is supplied.
This setting ensures that no improper RF signal due to the missing external reference
signal is output and used for measurements.
In addition to the error message "Ext Ref missing", the instrument generates the mes-
sage "RF output deactivated'.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
--
Example:
ROSC:SOUR EXT
Selects the external source. The reference must be input at the
REF IN input.
Example:
ROSC:EXT:RFOF:STAT ON
In case of a missing external signal, no RF signal is output.
Manual operation:
See
"Deactivate RF Output (if external reference is missing)"
[:SOURce]:ROSCillator:EXTernal:SBANdwidth
<SBandwidth>
Sets the synchronization bandwidth for an external reference signal.
SOURce Subsystem