Remote Control Commands
R&S
®
SMF100A
597
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
:TRIGger<hw>[:SWEep][:IMMediate]
Starts all sweeps which are activated for the respective path. The command starts all
sweeps which are activated.
The sweep to be executed depends on the respective MODE setting
(
:SOUR:SWEep:POW|FREQ:MODE
and
:SOUR:LFO:SWEep[:FREQ]:MODE
).
The command corresponds to the manual-control command "Execute Trigger".
Example:
TRIG
starts all active sweeps.
Usage:
Event
Manual operation:
See
" Execute Single Sweep - Frequency Sweep "
:TRIGger<hw>[:IMMediate]
The command immediately starts the activated sweep.
The command performs a single sweep and therefore applies to sweep mode AUTO
with sweep source SINGle. Use the commands
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING
, and
SOUR:SWE:FREQ|POW:MODE
, or
SOUR:LFO:SWE:[FREQ:]MODE
to set the respective sinlge sweep. You can alterna-
tively use an
IMMediate
command instead of the respective
SWEep:
[FREQ:]|POW:EXECute
command.
Example:
TRIG
starts all active sweeps.
Usage:
Event
Manual operation:
See
7.18 UNIT Subsystem
The
UNIT
subsystem contains the commands specifying which units are valid if no unit
is indicated in a command. These settings are valid for the entire instrument.
:UNIT:ANGLe
<Angle>
Defines the default unit for the phase modulation angle. It is not valid for commands
which determine angle values, e.g. RF phase. It does not influence the manual control
parameter unit and the display.
Parameters:
<Angle>
DEGRee | RADian
*RST:
RADian
UNIT Subsystem