Instrument Function
R&S
®
SMF100A
188
Operating Manual 1167.2319.02 ─ 12
Execute Single Sweep - Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single"
is selected.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:EXEC
Remote command:
[:SOURce<hw>]:SWEep:POWer:EXECute
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>[:SWEep][:IMMediate]
Start Level - Level Sweep
Sets the start level.
Remote command:
Stop Level - Level Sweep
Sets the stop level.
Remote command:
Current Level - Level Sweep
Displays the current level.
If "Step" is set, the level for the next level step of the sweep is entered here.
Remote command:
Shape - RF Level Sweep
Selects the cycle mode for a sweep sequence (shape).
"Sawtooth"
One sweep runs from the start level to the stop level. The subsequent
sweep starts at the start level again, i.e. the shape of sweep
sequence resembles a sawtooth.
"Triangle"
One sweep runs from start to stop level and back, i.e. the shape of
the sweep resembles a triangle. Each subsequent sweep starts at the
start level again.
Remote command:
[:SOURce<hw>]:SWEep:POWer:SHAPe
Retrace - RF Level Sweep
Activates that the signal changes to the start level value while it is waiting for the next
trigger event. It allows you to shift down the power during the waiting period.
RF Level