Remote Control Commands
R&S
®
SMF100A
532
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
[:SOURce<hw>]:POWer:STARt
<Start>
Sets the start level for the RF sweep.
Note:
You can select any level within the setting range. The range is defined by this
start value and the
A defined offset (
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
affects the level values according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
<Start>
float
Determines the first level value of the sweep setting range.
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
-30
Example:
POW:STAR -20 dBm
sets the start level for the level sweep to -15 dBm for RF output
A.
Manual operation:
See
[:SOURce<hw>]:POWer:STEP[:INCRement]
<Increment>
Sets the step width for
To adjust the level step by step with this step size, use the
POW:UP
and
POW:UP
com-
mands.
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the level accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
float
Range:
full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST:
1
Example:
POW:STEP 2
sets the step width for entering the RF level to 2 dB.
Manual operation:
See
SOURce Subsystem