Remote Control Commands
R&S
®
SMF100A
524
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:POWer:ALC:SONCe
Temporarily activates level control for correction purposes.
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:SONC
level control is performed once only.
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:POWer:ALC:SOURce
<Source>
The command selects the detector for level control.
Parameters:
<Source>
INTernal | DIODe | PMETer
INTernal
The internal detector is switched on.
DIODe
A diode detector can be connected to the EXT ALC input.
PMETer
A power meter can be connected to the EXT ALC input.
*RST:
INTernal
Example:
POW:ALC:SOUR INT
sets the internal detector for level control.
Manual operation:
See
" Source - External Level Control "
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
Activates automatic level control.
Parameters:
<State>
0 | OFF | 1 | ON
ON
Internal level control is permanently activated. This provides the
highest level accuracy.
OFF
Internal level control is deactivated,
Sample & Hold
mode is
activated.
*RST:
1
Example:
POW:ALC ON
activates internal level control.
Manual operation:
See