Instrument Function
R&S
®
SMF100A
195
Operating Manual 1167.2319.02 ─ 12
"PMeter"
A power meter can be connected to the EXT ALC input. The RF level
is controlled by the external voltage provided by the power meter.
The reference voltage for the desired level at the RF output of the
generator has to be entered below. Please see the specification of
the power meter for details.
Remote command:
[:SOURce<hw>]:POWer:ALC:SOURce
Power Meter Type - External Level Control
Selects the power meter that is used for external level control. This feature is only
available if PMeter is selected in the Source field.
Remote command:
[:SOURce<hw>]:POWer:ALC:PMETer:TYPE
Reference Voltage - External Level Control
Sets the reference voltage for the desired level at the RF output of the generator. See
the specification of the external source (diode detector or power meter) for details.
Remote command:
[:SOURce<hw>]:POWer:ALC:REFerence
5.4.6 Power Sensors
The R&S
SMF supports R&S NRP-Z power sensors to measure the power of the out-
put signal in the RF path or any freely selectable source. The generator performs up to
four power measurements, with the sensors either directly connected, or via the R&S
NRP-Z5 USB Sensor Hub.
The R&S NRP-Z5 USB Sensor Hub (high-speed USB 2.0) can host up to 4 R&S NRP-
Z sensors and provides simultaneous internal and external triggering of all connected
sensors. You can directly connect the R&S NRP-Z Sensors to the standard NRP sen-
sor connectors of the hub. The hub is connected to the R&S instrument either with the
R&S NRP-Z2 extension cable, or via USB using with the adapter cable R&S NRP-Z4.
See also the R&S website
Voltmeters > R&S NRP-Z Sensors" for information on the sensor hub and the available
accessories.
To connect an R&S NRP-Z sensor directly, the R&S
SMF provides the SENSOR con-
nector. A sensor connected there is always assigned as Sensor 1. Alternatively, you
can connect a sensor at a USB interface, using one of the USB adapters R&S NRP-Z3
or R&S NRP-Z4.
The instrument automatically detects a connected R&S NRP-Z power sensor and indi-
cates it in the
dialog. In addition, you find device specific infor-
mation on the connected sensor in
Chapter 5.2.3.5, "NRP-Z Info/Update"
on page 107. For information on the scope of your power sensor refer to the manual of
your R&S NRP-Zxx power sensor.
RF Level