Remote Control Commands
R&S
®
SMF100A
529
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:POWer:MODE
<Mode>
Sets the instrument operating mode and therefore also the commands used to set the
output level.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW
and
FIXed
are synonyms. To set the output level value, use
the command
SWEep
Operates in power sweep mode.
Set the range and current level with the commands
.
*RST:
CW
Example:
POW:MODE SWEep
selects the SWEep mode using the
POW:STAR; POW:STOP; POW:MAN
settings.
Manual operation:
See
[:SOURce<hw>]:POWer:POWer
<Power>
Sets the RF level of the RF output connector.
The level entered with this command corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note:
The SCPI command
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
sets the level of the "Level" display, i.e. the level containing offset.
Parameters:
<Power>
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The values are given in the data sheet.
Increment: 0.01
*RST:
-30
Example:
SOUR:POW:POW 15
sets the RF level at output to 15 dBm.
Manual operation:
See
[:SOURce<hw>]:POWer:RESolution
<Resolution>
The command selects the resolution for the level settings.
SOURce Subsystem