the two LEDs glow continuously green, and the AutoRAE 2 Controller’s display shows
the word “Ready” next to the instrument name without highlighting. (
Note:
If there is a
sensor mismatch, the name will be highlighted in yellow.) This indicates that you may
now perform a bump test or calibration.
If the instrument fails to warm up, the Cradle’s Bump and Cal LED lights will blink red
in alternation and the buzzer will alarm. The AutoRAE 2 Controller’s display highlights
the instrument in red with the word “Error.” Remove the monitor from the cradle and
refer to the information on the instrument display.
19.7. Testing
When the AutoRAE 2 Controller is turned on, it performs a self-test, SD card checks
(whether an SD card is present, whether it is full or nearing full capacity, etc.), a test of
each attached AutoRAE 2 Cradle, and then a test of any instruments that are in the
cradles. Bump testing and calibration can only take place on an instrument if its tests are
passed, including compatibility between the gas settings in the AutoRAE 2 Controller
and the instrument.
19.7.1. Compatibility Testing
The AutoRAE 2 Controller checks that the gas settings programmed into the AutoRAE 2
Controller match the settings in the instrument for each sensor and its calibration gas. If
there is a mismatch, the “Status” column on the display shows “Warning.” Press [N/-] to
scroll to the instrument in the list that you want to read details on. Press [MODE] to get
more info:
Summary of Contents for AutoRAE 2
Page 1: ......
Page 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Page 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Page 76: ......
Page 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Page 121: ......