21.1. Gas Inlet Settings
“Gas Inlet Information” tells the AutoRAE 2 what kind of gas is supplied to each gas
inlet. The “Gas Inlet Information” section includes configuration parameter settings for
the five gas inlets on the AutoRAE 2 Controller. For each “Gas Inlet,” you can view and
set the gas type, concentration, concentration unit, purge time, and soak time. You can
modify these values and upload them to your AutoRAE 2 or download the values
currently programmed into the AutoRAE 2 to ProRAE Studio II.
There are five gas configurations (Gas Inlet 1 through Gas Inlet 5), and each can be
customized to suit the number and types of gases in each gas inlet.
Select a Gas Inlet under “Gas Inlet Information”:
The window now shows the selected Gas Inlet (Gas Inlet 1 shown), its Gas Number
(number of gases in the gas mix, if the cylinder contains more than one gas; single gas
shown), Gas Lot Number, Expiration Date, and data that includes Gas Index, Gas Name,
Concentration, Concentration Unit, Purge Time (Sec.), and Soak Time (Sec.).
21.2. Configuring A Gas Inlet
Each Gas Inlet section is designed so that you can define individual gases or gas
combinations to match the cylinders you are planning to use with an AutoRAE 2
Controller-based system. Besides the number of gases, you can change each one’s Gas
Index (and Gas Name), Concentration, Concentration Unit, Purge Time, and Soak Time.
In addition, you can change the Gas Lot Number and expiration date for each cylinder.
21.2.1. Gas Number
This is the number of gases (up to eight) present in a single Gas Inlet. Click the up arrow
to increase the number or the down arrow to decrease the number.
Note:
The number can
Summary of Contents for AutoRAE 2
Page 1: ......
Page 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Page 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Page 76: ......
Page 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Page 121: ......