23. Transferring AutoRAE 2 Controller Data To A Computer
Bump and calibration data is collected each time an instrument is bump tested or
calibrated. The AutoRAE 2 Controller collects this data and stores it on an SD Card,
providing a convenient means of storage and easy data transfer.
Follow this procedure for reading calibration and bump testing data from an AutoRAE 2
controller:
1.
Connect a USB cable to the AutoRAE 2 Controller and to a PC running ProRAE
Studio II software
2.
Make sure the AutoRAE 2 Controller has power and is turned on.
3.
Start ProRAE Studio II software on the PC.
Note:
You can get reports with any level of access privileges.
4.
Click “Detect the instruments automatically” (the magnifying glass icon with the
letter “A” in it). After a few seconds, the AutoRAE 2 Controller is found and it is
shown, along with its firmware version, serial number, COM port:
5.
Click “Select.”