9. Bump And Cal Error And Status Messages
The LEDs labeled “Bump” and “Cal” above the two keys on the AutoRAE 2 Cradle
provide information about status during bump and calibration testing. The following table
explains the various messages:
Status
Bump LED
Cal LED
User Action
Warm-up in
progress
Orange
blinking in
alternation
with Cal LED
Orange
blinking in
alternation with
Bump LED
Pressing keys has no effect.
Warm-up error
Red
blinking in
alternation
with Cal LED
Red
blinking in
alternation with
Bump LED
Remove the instrument from the
cradle and take the action
indicated on the instrument’s
display.
Warm-up
completed
successfully
Green (solid)
Green (solid)
Press Bump or Cal key to
perform a bump test or
calibration.
Bump test in
progress
Green
(blinking)
Off
Do not remove the instrument, or
the process will be interrupted.
Pressing keys has no effect.
Bump test
completed
successfully
Green (solid)
Off
The bump test result (pass) has
been logged. You may remove
the instrument from the Cradle
for use or leave it on the Cradle
for the next AutoRAE 2
operation or to charge its battery.