Status
Bump LED
Cal LED
User Action
Bump test failed
Red (blinking
slowly)
Off
The bump test result (fail) has
been logged.
Calibration in
progress
Off
Green
(blinking)
Do not remove the instrument, or
the process will be interrupted.
Pressing keys has no effect.
Calibration
completed
successfully
Off
Green (solid)
The result (pass) has been
logged. You may remove the
instrument from the Cradle for
use or leave it on the Cradle for
the next AutoRAE 2 operation or
to charge its battery.
Calibration failed Off
Red (blinking
slowly)
The result has been logged in the
instrument; you may remove it
from the Cradle and read its
display for an error code. If a
Controller is used, the
Controller’s display shows the
error code.
Sleep mode
Orange (solid)
Orange (solid)
Charging continues when in
sleep mode. Press either key to
wake up the instrument.
Monitor not
connected /
system idle
Off
Off
Check to make sure the monitor
is properly installed in the cradle
and the capture mechanism is
fully engaged.
Check to make sure the monitor
is in AutoRAE 2
Communications Mode or turned
off.
Cradle error
Red (blinking
slowly at the
same time as
the Cal LED)
Red (blinking
slowly at the
same time as
the Bump LED)
Take the action indicated in the
Controller’s display, if used.
Otherwise, contact RAE Systems
Technical Support.
Summary of Contents for AutoRAE 2
Page 1: ......
Page 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Page 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Page 76: ......
Page 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Page 121: ......