If the system’s initial checkout passes, then this screen is displayed:
If either or both mechanical tests fail, then a red square with an “X” in it indicates the
problem:
If the Controller Valve & Manifold Check fails, you should turn off the system and
contact RAE Systems Technical Support.
If the Terminal Adapter Check fails, make sure the Terminal Adapter is connected to the
last AutoRAE 2 Cradle, and that all of the AutoRAE Cradles in the system are firmly
connected. Try restarting the system. If everything is connected but the test fails again,
contact RAE Systems Technical Support.