Summary Of Differences Between An AutoRAE 2 Controller-Based System
And A Stand-alone Cradle
Controller-Based System
Standalone Cradle
Monitors calibrated simultaneously Up to 10
1
Number of gas inlets (distinct
calibration gas cylinders)
5 dedicated gas inlets plus
fresh air with dedicated
exhaust port on the
Controller
2 dedicated gas inlets
plus fresh air with
dedicated exhaust port
Display
5.7" TFT (320 x 240)
graphical backlit color
LCD on the Cont
2 seven-segment LED
displays on each Cradle
2 seven-segment LED
displays
Buttons
3 buttons ([Mode], [Y/+],
and [N/-]) on the
Cont 2 buttons
([Bump] and [Cal]) on
each Cradle
2 buttons ([Bump] and
[Cal])
Power supply
12V, 7.5A output
Charges up to 10
instruments at a time
12V, 1.25A output
Charges one
instrument at a time
Printer support
Direct printing on
Serial (RS-232)
printers
Printing
Automatic
Built-in pump
Built-in pump
(500 ml/min) in the
Controller
MultiRAE Cradle relies
on MultiRAE’s pump
to pull in air.
ToxiRAE Pro Cradle
has 300cc/min internal
pump
AutoRAE Cradle For
Handheld PID relies on
the instrument’s pump
to pull in air.
Data storage
Standard 2 GB SD card
with security lock on the
Controller
None. Data stored only
on the PC
Networking
RJ-45 10/100 Base-T port
on the Controller
Wireless connectivity via
external WiFi adapter
None
Summary of Contents for AutoRAE 2
Page 1: ......
Page 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Page 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Page 76: ......
Page 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Page 121: ......