16. Operation of an AutoRAE 2 Controller-based System
Deploying AutoRAE 2 Cradles with the AutoRAE 2 Controller significantly enhances
the system’s capabilities compared to those of a standalone cradle. An AutoRAE 2
Controller-based system can charge, test, and calibrate up to 10 instruments at the same
time using up to 5 distinct gas inputs. The system supports USB printers (with HP Printer
Command Language 5 or 5E) for direct printing, and has a standard SD card on which
data are stored.
When one or more AutoRAE 2 Cradles are attached to the AutoRAE 2 Controller, the
AutoRAE 2 controller acts as the “command center” for the system. The Controller
powers the entire system, manages all the configuration settings, and its built-in pump
and valves control the gas flow.
Note:
An AutoRAE 2 Terminal Adapter must be attached to the rightmost cradle in order
for an AutoRAE 2 Controller-based system to work.
IMPORTANT!
Before using the AutoRAE 2 Controller, it must be configured using ProRAE Studio II
software to set the gas types and concentrations, as well as the time and date. See page 88
for details.
Summary of Contents for AutoRAE 2
Page 1: ......
Page 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Page 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Page 76: ......
Page 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Page 121: ......