12.4. Concentration [Value]
You can set the concentration by double-clicking in the respective gas concentration box
and then typing in the concentration value.
12.5. Concentration Unit
When you pull down the Concentration Unit menu, select the desired gas concentration
units (there are other types of units).
ppm
%
ppb
mg
ug
%LEL
%VOL
%CH
4
12.6. Purge Time (Sec.)
Type to set the number of seconds for the system to purge with fresh air after performing
a bump test or calibration.
12.7. Soak Time (Sec.)
Type to set the number of seconds for the system to allow the sensor to be pre-exposed to
calibration gas before bumping or calibrating.