21. Programming An AutoRAE 2 Controller-based System
on the Computer
When it comes to an AutoRAE 2 Controller-based system configuration, some
parameters, such as date and time, can be configured either on the PC or directly on the
Controller screen. Other parameters, such as the AutoRAE 2 system password can only
be configured directly on the Controller, whereas gas configurations (gas inlet settings),
can be only configured on a PC. Firmware updates for both the Controller and all the
Cradles connected to it can also only be done on a PC.
To program an AutoRAE 2 Controller-based system on a PC, you need ProRAE Studio II
Instrument Configuration and Data Management Software, the AutoRAE 2 Controller-
based system connected to a power source, and a USB PC communications cable.
1.
Connect a USB cable between a PC with ProRAE Studio II and the AutoRAE 2
Controller.
2.
Turn on the AutoRAE 2 Controller (with AC adapter connected and plugged in,
power switch toggled so red LED light is glowing).
3.
Start ProRAE Studio II software on the PC.
4.
Select “Administrator” and input the password (the default is “rae”).
5.
Click “Detect the instruments automatically” (the magnifying glass icon with the
letter “A” in it). After a few seconds, the AutoRAE 2 Controller is found and it is
shown, along with its firmware version, serial number, COM port:
6.
Click “Select.”
7.
Click “Setup.”
Summary of Contents for AutoRAE 2
Page 1: ......
Page 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Page 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Page 76: ......
Page 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Page 121: ......