Each adapter comes with a sticker for the inside of the cradle. It is especially
recommended that you install these stickers, particularly if both types of adapters are
being used in multiple cradles connected to an AutoRAE 2 Controller.
6.6.2. Placing A ToxiRAE Pro Monitor In The Cradle
1.
Make sure the correct adapter is located in the cradle’s capture mechanism.
2.
Make sure the external filter on the instrument is not dirty or clogged and screwed
onto the instrument inlet tightly.
3.
Make sure the monitor is either turned off or is in AutoRAE 2 Communications Mode.
4.
Place the instrument into the cradle face-down, making check that it is aligned
correctly with the contacts on the AutoRAE 2 Cradle’s charging port.
5.
Press in on the capture mechanism to lock the ToxiRAE Pro in place.
Note:
There is no need to remove the external filter, rubber boot, belt clip or wrist strap
from the monitor to use it with the AutoRAE 2.
Place self-adhesive
sticker here