882 Video Test Instrument User Guide
(Rev A.35)
437
LSGX:PLAT
LipSync Generator Data Stream Order
Class
LipSync Generator/Analyzer
Description
Determines which data stream (audio or video) is transferred (sent) first for lipsync test
image.
Command syntax
LSGX:PLAT
boolean
boolean
0 = video is transferred first (indicating that video takes longer to process–audio is
delayed).
1 = audio is transferred first (indicating that audio takes longer to process–video is
delayed).
Query syntax
LSGX:PLAT?
Query response
0 or 1
Example
LSGX:PLAT 1
// Audio is sent first for lipsync test image
LSGX:TLAT 0.5
// Latency between audio and video in lipsync test image
LSGX:LSGU
// Update hardware to new settings
Summary of Contents for 881
Page 1: ...881 882 Video Test Instrument User Guide 882E for HDMI ...
Page 12: ...10 Contents ...
Page 50: ...38 Chapter 1 Getting Started ...
Page 84: ...72 Chapter 2 Testing Video Displays ...
Page 99: ...882 Video Test Instrument User Guide Rev A 35 87 ...
Page 102: ...90 Chapter 3 Administrative Tasks ...
Page 107: ...882 Video Test Instrument User Guide Rev A 35 95 ...
Page 216: ...204 Chapter 5 Working with Formats ...
Page 248: ...236 Chapter 7 Working with Images ...
Page 264: ...252 Chapter 8 Working with Test Sequences ...
Page 382: ...40 Chapter 11 Testing EDID for HDMI 2 Expand the EDID Block in the navigation view ...
Page 446: ...104 Chapter 12 CEC Interactive Troubleshooting Environment ITE ...
Page 493: ...882 Video Test Instrument User Guide Rev A 35 151 ...
Page 591: ...882 Video Test Instrument User Guide Rev A 35 213 Commands by name ...
Page 687: ...882 Video Test Instrument User Guide Rev A 35 309 Example DVIC 23 FMTU ...
Page 1018: ...580 Appendix B Image Reference The EMITest5 image is shown below ...
Page 1094: ...656 Appendix B Image Reference ...
Page 1124: ...686 Appendix C Error Messages ...
Page 1140: ...702 Appendix D Format Reference ...