Appendix B
Gemini VII
B-14
239-42828-01 - Nov 2013
4039-
The isotherm does not meet the constraints of the Cheng/Yang assumption.
Cheng/Yang correction will not be applied.
4040-
The value of Qm computed from the Langmuir equation is too low. The pore
size will not be computed for all data points.
4041-
Cheng/Yang correction is inappropriate for some
P/Po
.
Cause:
The Cheng/Yang correction to the Horvath-Kawazoe equation
requires the value of the monolayer volume as calculated from the
Langmuir equation. The isotherm does not correlate to the Langmuir
assumption with a coefficient of 0.98 or more. The correction is not
applicable to this isotherm or to the range of the data points selected.
Action A:
The analysis will be performed without the Cheng/Yang correction.
Deselect Apply Cheng/Yang correction on the Horvath-Kawazoe
Report Options window to prevent this message from appearing on
future reports.
Action
B:
Generate the Langmuir report for the same data points selected for the
Horvath-Kawazoe report. If the Langmuir correlation coefficient can
be brought above 0.98 by removing some points at high relative
pressure, remove them and reproduce the Horvath-Kawazoe reports.
Cause:
The Cheng/Yang correction to the Horvath-Kawazoe equation
requires the value of the monolayer volume as calculated from the
Langmuir equation. The computed value is less than the volume
adsorbed at the largest relative pressure included in the analysis. The
correction is not applicable to this isotherm or to the range of the data
points selected.
Action:
The analysis will be performed and the Cheng/Yang correction will
be applied to all points with a volume adsorbed less than the value of
Vm. The pore size will not be calculated for data points with an
invalid volume adsorbed. Deselect Apply Cheng/Yang correction on
the Horvath-Kawazoe Report Options window to clear this message.
Cause:
The Cheng/Yang correction is usually inappropriate for any P/Po
above the isotherm knee. In some instances, the computed pore sizes
may decrease above the knee. While it is possible to include these
relative pressures (usually above 0.1 P/Po) in the analysis, the
computed pore sizes for these pressures are usually meaningless.
Action:
Change the data points selected for the Horvath-Kawazoe report to
include only relative pressures at or below the knee of the isotherm or
change the Horvath-Kawazoe report options so that the Cheng/Yang
correction is not applied.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...