8 E
RROR
M
ESSAGES
Program error messages are listed numerically. If the
Action
response indicates to contact a
Micromeritics service representative, record the error message, then make backup copies of any files
involved in the operation.
4407
Error searching for installed Smart VacPreps. The registry could not be read..
Cause:
The application was not installed properly.
Action:
Reinstall the application. If the problem persists, contact your Micromeritics service rep-
resentative.
4408
The %s in %s already has Smart VacPrep S/N %d. The Smart VacPrep
must be removed from the %s before it can be added.
Cause:
The Smart VacPrep was already installed for another application / unit.
Action:
Remove the Smart VacPrep from the installed unit before adding it to the preferred unit.
4409
A free IP address on the same subnet as %s could not be found."
Cause:
All IP addresses on the network for the Ethernet card specified during installation are in
use by other Micromeritics applications on this computer.
Action:
Uninstall unused Micromeritics applications.
Action:
Configure a different Ethernet card for use by the application using the application
installer.
4410
The .INI file could not be updated with configuration for Smart
VacPrep S/N %s."
Cause:
The application .INI file is opened by another application and could not be updated.
Action:
Close all open applications and add the Smart VacPrep again using the Smart VacPrep
menu.
4411
Error dosing.
Cause:
The backfill timed out.
Action:
Ensure there is gas available and the pressure regulator is set to the appropriate pressure.
Also ensure that the gas supply regulator shut-off valve is open.
8 Error Messages
Smart VacPrep Operator Manual
8 - 1
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...