Appendix B
Gemini VII
B-20
239-42828-01 - Nov 2013
4073-
The analysis gas [name] does not match the model gas [name].
4074-
The analysis temperature [nn] does not match the model temperature [nn].
4075-
The models cannot be located in the models folder. Reinstall the software.
4077-
Cannot get surface area for: [file name]
.
4078-
Slope and Y-Intercept cannot be determined from the selected points.
4079-
Points found with negative pressure values will not appear on the graph.
Cause:
The model assumes a specific gas and the sample file uses a different
one.
Action:
Select a model that assumes the same gas.
Cause:
The temperature for the selected model did not match the analysis
temperature.
Action:
Select a different model.
Cause:
The models could not be located. They may have been inadvertently
deleted or moved.
Action:
Reinstall the software.
Cause:
The Isotherm report for the named overlay file has
Per gram
selected
for the
Volume Adsorbed
and the Isotherm report for the primary file
has a surface area option selected for the
Volume Adsorbed
.
Action A:
Edit the Isotherm report for the named overlay file and select a
surface area option for
Volume Adsorbed
.
Action B:
Click
Overlays
on the Report options window of the primary file and
remove the named overlay file from the list.
Cause:
The Langmuir report cannot be generated from the selected points.
Action:
Edit the calculation assignments in the Langmuir column on the
Collected Data window.
Cause:
Collected data contains negative pressure values.
Action:
Negative pressure points are automatically removed from isotherm
graphs. To display negative pressure values appear on the graph:
Open the sample file and click the
Collected Data
tab. Deselect
Use
calculation assignments
and then select the points with negative
pressures as Outliers.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...