Gemini VII
Connecting Gases
239-42828-01 - Dec 2012
9-19
2.
Both gauges should read at or near zero. If not, open the regulator gas shut-off valve to release
gas. It is not necessary to disconnect the gas line from the regulator or the instrument.
3.
Use an appropriate wrench to loosen the nut at the regulator/gas bottle connection then remove
the regulator from the bottle.
4.
Replace the protective cap on the depleted bottle. Disconnect the retaining strap and remove the
bottle from its current location.
Connecting a Replacement Gas Bottle
Move the replacement bottle close to the instrument and tether it into place. It is not necessary to dis-
connect the gas line from the regulator or the instrument.
1.
Use an appropriate cylinder wrench to remove the protective cap from the replacement bottle.
When connecting hazardous gases, ensure there is proper ventilation and always
follow the safety procedures established for your lab.
A power failure or loss of cryogen can result in dangerous pressures in the sample
tube. The analyzer uses pressure relief valves to vent this pressure into the instrument
cabinet and return the instrument to a safe condition. When using toxic or flammable
gases, additional venting of the cabinet may be required.
Regulator gas shut-
off valve
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...