Gemini VII
Performing a Reference Analysis
239-42828-01 - Dec 2012
9-29
Analyzing Reference Material
1.
Select
Unit [n] > Start Analysis
.
2.
On the
Analysis
screen, click
Browse
to select the test file created in the previous steps. Click
Open
.
3.
Click
Edit
to make and save modifications or replace all parameters in the open file using the
Replace All
button to select another file. Click
Save
and
Close
when done.
4.
Click
Report after analysis
to automatically generate reports to the specified destination when
the analysis is complete. On the
Report Setting
s window, select the report destination. Click
OK.
5.
Click the
Start
button. Graphs are displayed as the analyses are performed.
When the analyses are complete, compare the BET Surface Area shown on the Summary Report
with the BET Surface Area shown on the reference material bottle. The values should match
within the tolerance level shown on the bottle.
If the BET Surface Area values do not match, refer to the following causes and actions.
Cause
Action
The sample was not degassed properly.
Degas the sample again.
The gas lines are not clean.
Perform the procedure for cleaning and verifying
gas lines, then try again.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...