Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-17
4053- At least two data points must be selected for t-Plot calculations.
4054-
Fewer than two data points are inside the fitted thickness range.
4055-
A positive BET surface area was not calculated.
4056- A positive Langmuir surface area was not calculated. Please check your Lang-
muir report.
4057-
At least two data points are needed for Freundlich calculations.
Cause:
At least two points are required for the t-Plot calculations.
Action:
Edit the calculation assignments for the t-Plot report.
Cause:
At least two points must be within the fitted thickness range for the t-
Plot calculations.
Action A:
Edit the calculation assignments for the t-Plot report.
Action B:
Edit the fitted thickness range in the t-Plot report editor.
Cause:
Fewer than two points were assigned to the requested surface area
calculation in the collected data table.
Action A:
Assign more points to the surface area calculation.
Action B:
Select a different surface area in the t-Plot report editor.
Cause:
Fewer than two points were assigned to the requested surface area
calculation in the collected data table.
Action A:
Assign more points to the surface area calculation.
Action B:
Select a different surface area in the t-Plot report editor.
Cause:
Less than two data points have been selected for the Freundlich
report; at least two are required.
Action:
Select Freundlich points on the Collected Data window. If calculation
assignments are not being used, edit the Freundlich Report options,
Absolute pressure range
in the sample file.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...