Serial write queue full.
6026-
Measure Po for each point is not supported by this model. Choose another
sample file for analysis or change the Po option in the analysis conditions.
6027-
Analysis: Po dosing failure
.
6028-
This instrument does not provide measured Po analyses.
6029-
Instrument is not completely calibrated.
Cause:
There was a fatal error in communication between the analyzer and
the software. All displays for that instrument will be closed.
Action:
Ensure the Ethernet cable is securely connected. Exit the application
and then restart it. Contact a Micromeritics service representative if
this error message continues.
Cause:
Measure Po for each point
was selected as the saturation method.
This option is only available for the Gemini 2390
p
or 2390
t
.
Action:
Edit the Po options in the analysis conditions and select
Entered
or
Most recent measured
as the saturation pressure method.
Cause:
The Dewar does not contain enough cryogen to take an accurate
reading in the Po tube.
Action:
Refill the Dewar with cryogen and restart the analysis.
Cause:
An attempt was made to start an analysis on the Gemini 2390
a
using
a sample file in which
Measure Po for each point
was selected as the
saturation method. This feature is available only for the Gemini
2390
p
and 2390
t.
Action A:
Cancel the analysis, open the sample file, and choose
Entered
or
Most
recent
measured
for the Po value. Restart the analysis.
Action B:
Select a different file and continue with the analysis.
Action C:
If a Gemini 2390
p
or 2390
t
is attached, select
Start Analysis
for that
unit.
Cause:
The instrument in use is missing calibrations.
Action:
Select
Unit > Calibration > Load from File
and select a file
containing calibration data. Contact a Micromeritics service
representative if this error message continues.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...