Performing a Blank Analysis for Diagnostic Purposes
Gemini VII
9-26
239-42828-01 - Dec 2012
Performing a Blank Analysis for Diagnostic Purposes
A blank analysis is a means of pinpointing operational problems in the analyzer. Leaks, as well as elec-
tronic problems, will result in an unsatisfactory blank analysis. This test should be performed prior to
detailed troubleshooting. The data produced during the blank analysis can be very helpful in identify-
ing the cause of a problem and should be performed before contacting your Micromeritics Service
Representative.
Helium is required for performing a blank analysis since a measured free space is required. If helium is
unavailable, contact a Micromeritics Service Representative for an alternative method of performing
this procedure.
1.
Install empty, straight-wall sample tubes (free of cracks or chips) of the same size in the sample
and balance ports. Ensure the O-rings are present and in good condition.
2.
Place the Dewar support on the elevator, then place a Dewar of water at ambient temperature on
the Dewar support.
A Dewar support is not necessary for the larger Dewar used with the Gemini VII 2390t. Place the
Dewar directly on the elevator.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...