Appendix B
Gemini VII
B-12
239-42828-01 - Nov 2013
4029-
At least two fitted data points are needed for Alpha-S calculations.
4030-
Preparations failed in primary data.
4031-
Not enough points with a relative pressure in the range [n,n]
4033-
Not enough points to generate Dubinin Tabular Report.
4034-
Fewer than 2 points available for Dubinin calculations.
4035-
Cannot calculate optimized Astakhov exponent: Not enough points with a rel-
ative pressure in the range [(pressure), (pressure)].
Cause:
Fewer than two data points fall within the selected Alpha-S range.
Action:
Edit the calculation assignments or the fitted Alpha-S range or use a
different reference curve.
Cause:
Appropriate data were not available to generate the report.
Action:
This message was preceded by a different error message. Refer to the
cause/action of the preceding message.
Cause:
Fewer than two data points selected for the Dubinin report fall within
the selected relative pressure range.
Action:
Edit the calculation assignments or the fitted relative pressure range.
Cause:
There are fewer than two valid data points available for the Dubinin
tabular reports.
Action:
Examine the calculation assignments in the Collected Data window
of the sample file. At least two micropore pressures must be selected
for inclusion in the Dubinin report.
Cause:
There are fewer than two valid data points available for Dubinin
reports in one of the sample files selected for overlaying.
Action:
Examine the calculation assignments in the Collected Data window
of the sample file. At least two micropore pressures must be selected
for inclusion in the Dubinin report.
Cause:
There are fewer than two valid data points in the relative pressure
range specified. Astakhov reports will not be produced.
Action:
Examine the calculation assignments in the Collected Data window
for the sample file. At two pressures in the given range must be
selected for inclusion in the Dubinin report.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...