Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-7
2499-
Sequence number must contain at least 3 digits.
2500 Series
2500-
All sample file names that can be created using the sequence number pattern
already exist. You may want to modify the next sequence number.
2505-
Error logger cannot be initialized. Error code [n]. Program will exit.
2506-
Sample file [file name] has a 'No Analysis' status and cannot be used for this
operation.
2508-
[sample file] Overlay file [name] was not found. It will not be included in the
reports.
2509-
[sample file] Error opening file [name] [error]. Reports cannot be produced.
Cause:
An attempt was made to enter a sequence number that did not contain
at least three digits.
Action:
Enter a sequence number that contains at least three digits.
Cause:
No more sample information files can be created using the currently
entered file name sequence number.
Action: Select
Options > Sample defaults
and enter another sequence
number.
Cause:
An internal processing error has occurred.
Action:
Contact a Micromeritics service representative if this error message
continues.
Cause:
The selected sample file does not have collected data and cannot be
used for operations, for example, reporting.
Action:
Enter the name of a file with a status of Complete, Analyzing or
Entered or select a sample file from the list box.
Cause:
The specified overlay file could not be found.
Action:
Ensure the file specified as an overlay exists.
Cause:
An error occurred while the program was opening a file necessary to
the report operation.
Action:
Use the name given in the error message to investigate. Contact a
Micromeritics service representative if this error message continues.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...