Schematic Shortcut Icon Available Options:
o
Cool the sample to less than 50 °C
o
Set heater power percent
Smart VacPrep Schematic Shortcuts Table (continued)
S
HOW
S
MART
V
AC
P
REP
S
TATUS
Smart VacPrep > Unit [n] > Show Status
The
Smart VacPrep Status
window allows the monitoring of degas operations for a manual degas
operation.
Field or Button
Description
Cancel
Cancels the current degas operation.
Check
Click to check the outgassing rate of the samples on the selected port.
The following process occurs:
l
The current degassing step is suspended.
l
The vacuum valves are closed and the vacuum level on the selected
ports is monitored.
l
The
Status
window indicates that the degassing operation is being
checked and displays the outgassing rate as it becomes available.
During the degas check, the
Check
button changes to
Continue
. When
Continue
is clicked, the valves open and the degassing operation
resumes.
If the outgassing rate indicated that the sample has been freed of
contaminants (minimal pressure increases), click
Skip
to advance to the
next state of the degassing operation. For example, if degassing is
checked after the setpoint is attained,
Skip
advances you to the ramping
stage.
Smart VacPrep Status Fields and Buttons Table
Show Smart VacPrep Status
Smart VacPrep Operator Manual
3 - 11
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...