Appendix B
Gemini VII
B-26
239-42828-01 - Nov 2013
6043-
Evacuation timeout.
6044-
Dosing timeout.
6045-
Elevator timeout.
6048-
Po fill timeout.
Cause:
The analyzer failed to reach a suitable vacuum level within the
specified evacuation time.
Action A:
Remove and reattach the sample and balance tubes, ensuring that both
are securely in the port.
Action B:
Verify that the vacuum pump is operating properly.
Action C:
If actions A and B don’t correct the problem, degas the sample again
and repeat the analysis.
Action D:
Contact a Micromeritics service representative if this error message
continues.
Cause A:
The adsorbate gas supply was depleted by the analyzer during
analysis or termination.
Action A:
Verify that the inlet valve on the gas bottle is open and that the gas
pressure regulator setting is between 15 and 18 psig (103.4 and 124.1
kPa). Check the adsorbate gas bottle to see if it is empty; replace if
necessary.
Cause B:
Sample has adsorbed the maximum amount of gas available.
Action B:
If additional data points are required, repeat the analysis with a
smaller amount of sample (to avoid exceeding 100 cm
3
STP total
volume adsorbed).
Cause:
The elevator was unable to reach the raised or lowered position.
Action:
Check for obstructions. Contact a Micromeritics service
representative if this error message continues.
Cause:
The adsorbate gas supply was depleted during and analysis or a Po
measurement.
Action:
Verify that the inlet valve on the gas bottle is open and that the gas
pressure regulator setting is between 15 and 18 psig (103.4 and 124.1
kPa). Check the adsorbate gas bottle to see if it is empty; replace if
necessary.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...