Appendix B
Gemini VII
B-22
239-42828-01 - Nov 2013
4405-
Fatal communications error with SmartPrep Unit [n] - S/N: [nn].
6000 Series
6000-
An error occurred while loading the application control information. Data entry
cannot be performed. [Code number]
6008-
The sample has an invalid status and cannot be used for this analysis.
6023-
Desorption points in this sample file cannot be processed by this instrument.
Continue with adsorption points?
6024-
Some selected reports are invalid for this instrument.
Cause:
There was a fatal error in the serial communications between the
application and the SmartPrep Instrument Controller. All displays for
that SmartPrep will be closed.
Action:
Ensure that the SmartPrep is properly chained to the computer on the
communications port configured in the Setup program. Stop and re-
start the application. Contact a Micromeritics service representative if
this error message continues.
Cause:
An error occurred accessing the control information disk file required
by this application.
Action:
The disk drive may have failed or be corrupt. Run diagnostics on the
disk drive.
Cause:
A sample file with a status other than
No Analysis
or
Prepared
was
selected.
Action:
Select a sample file with a status of
No Analysis
or
Prepared
.
Cause:
The selected sample file in this analysis has desorption points
selected in the pressure table. The analyzer (Gemini 2390
a
or 2390
p
)
cannot report desorption data.
Action A:
Click
Yes
to continue with the analysis and report adsorption data
only.
Click
No
to choose a different file, or to edit the current file.
Action B:
If using a Gemini 2390
t
, select
Start Analysis
for that unit.
Cause:
The requested report cannot be generated with the Gemini 2390
a
.
Action:
Edit the report selections on the Report Options window.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...