Gemini VII
Report Options Files
239-42828-01 - Dec 2012
5-19
Apply thermal
transpiration
correction checkbox
Select to correct the temperature-induced pressure difference between the
manifold and the chilled sample tube. This option is most significant for
pressures less than approximately 1.0 mmHg. Do not use filler rods in the
sample tube when applying correction for thermal transpiration.
Always use thermal transpiration when performing micropore analyses.
Refer to
for additional information on
thermal transpiration.
Inside diameter of
sample tube text box
Enabled when
Apply thermal transpiration correction
is selected. Enter
the inside diameter of the sample tube used in the analysis.
Selected Reports
list box
Select the checkbox to the left of the report names to include in the report.
Calculated pore width (w), pore radius (R) or pore diameter (D) for BJH
reports. Select
Options > Units
to specify default calculations.
Overlays button
Browse button
Close button
Edit button
Preview button
Replace button
Save button
Common Fields and Buttons - File Menu Options
.
Field or Button
Description (continued)
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...