Gemini VII
Preventive Maintenance
239-42828-01 - Dec 2012
9-5
Lubricating Elevator Drive Assembly
The elevator screw is lubricated before it leaves the factory and should not require lubricating. If the
elevator starts to vibrate or becomes noisy when traveling, contact a Micromeritics Service Represen-
tative for disposition.
Cleaning the Analyzer
The exterior casing of the analyzer may be cleaned using a clean cloth, dampened with isopropyl alco-
hol (IPA), a mild detergent, or a 3% hydrogen peroxide solution.
Do not use any type of abrasive
cleaner
.
Cleaning the Analysis Dewar
Ice and suspended frost particles may accumulate in the bottom of an analysis port Dewar. Particles or
deposits exceeding 1/4-in. (0.64-cm) in depth may jam between the bottom of the sample tubes and the
bottom of the Dewar, causing the Dewar not to raise fully. Accumulations of fine particles impede liq-
uid nitrogen circulation around the bottom of the sample tubes. This causes the sample temperature to
be slightly higher which, in turn, can cause pore volume measurement errors in those samples exhibit-
ing high isotherm slope above 0.97 relative pressure.
Accumulated ice is likely to melt and form a pool of water in the Dewar if all liquid nitrogen evapo-
rates. The water must be removed, otherwise it will solidify when liquid nitrogen is added and could
press on the bottom of the sample tube causing breakage.
To ensure problems do not develop due to ice accumulation, check the Dewar after each use. Clean the
Dewar on a weekly basis.
1.
Remove the sample tubes from the analysis ports.
2.
Go to
Unit [n] > Show Instrument Schematic
to display the instrument schematic, then go to
Unit [n] > Enable Manual Control
.
3.
Right-click on the elevator icon and select
Lower
to lower the elevator to its lowest position.
4.
Remove the Dewar and pour the liquid nitrogen from the Dewar into an appropriate cryogenic
container.
Do not allow liquid to penetrate the casing of the analyzer. Doing so could result in
damage to the unit.
When handling Dewars, be sure to observe the Dewar precautions outlined in
chapter 3.
Do not pour liquid nitrogen directly into a sink. Doing so may cause drain pipes to
burst.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...