Gemini VII
Sample Information Files
239-42828-01 - Dec 2012
5-9
Bar Code text box
Enter bar code reader information if a bar code reader is connected to the
computer’s USB port. If a bar code reader is not used, this alpha/numeric
field can be used to enter additional information about the sample; such as, a
sample lot number, sample ID, chain of custody, etc.
Mass group box
•
Enter
- when selected, enables the
Sample Mass
field. Enter a value for
the sample mass.
•
Calculate
- enables the
Empty tube
and
tube
fields. Enter the
values necessary to calculate the sample mass. Equation used to calculate
sample mass:
Mass
sample
= Mass
tube
– Mass
tube
•
Density
- value is used for the Calculated free space method only. Use
0.000
for a blank analysis.
Type of Data group
box
•
Automatically collected
- select if the type of data will be automatically
collected while an analysis is running.
•
Manually entered
- select to manually enter data. If
Manually entered
is selected, an
Entered
Data
tab displays at the top of the window.
User Parameters
group box
These fields are primarily used for the SPC (Statistical Process Control)
reporting to specify sample characteristics or its manufacturing process but
may be used for other data by entering specific analysis conditions or
sample criteria.
These parameters display on the Sample Description window, in
Reports >
SPC Report Options,
and the Summary Report.
Modify the default field labels (alpha/numeric) in
Options > Sample
Defaults
.
Comments text box
Enter comments about the sample or its analysis conditions. Comments will
appear in the report header.
Add Log Entry
button
Click to enter information to appear in the sample log report. Click the
button again to enter multiple log entries.
Replace All button
Click to select another file where the values will replace all values for the
Sample Information file. The original file will remain unchanged.
Field or Button
Description (continued)
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...