Preparing for Analysis
Gemini VII
3-18
239-42828-01 - Dec 2012
5.
Use rubber gloves to ensure no oils or residue are transferred to the clean tubes and filler rods and
remove the sample tubes and filler rods from the unit.
6.
Clean the interior of the sample tubes with the brush supplied with the system.
7.
Rinse the sample tubes and filler rods thoroughly with hot water. Then rinse again with isopropyl
alcohol or acetone. If isopropyl alcohol or acetone is not available, deionized water may be used.
8.
Stand the sample tubes on the sample tube rack and place the filler rods in a basket or in the rack.
Bake in a vacuum oven for two hours at 110 °C.
9.
Remove the sample tubes and filler rods from the oven and allow to cool.
10. Wipe a rubber stopper with a lint-free cloth.
11. Label the sample tube and stopper for identification.
Determining the Sample Mass
Analysis results are expressed in units of surface area per gram of sample; therefore, it is important the
true sample mass be known. The true mass is best calculated as:
• weigh the empty Sample Tube Set (sample tube and stopper or seal frit)
before degas
• weigh the Sample Tube Set with sample
before degas
and subtract the weight of the empty Sample
Tube Set
• weigh the Sample Tube Set with sample
after degas
and subtract the weight of the empty Sample
Tube Set
Do not insert the filler rods at this time. Filler rods are inserted just before the sample
tube is installed on the analysis port.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...