Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-27
6049-
Operator cancelled operation.
6050-
Valve failure.
6051-
Power failure.
6052-
Zero calibration failed.
6053-
Scale calibration failed.
6054-
Transducer match failed
6055-
Dosing servo calibration failed.
6056-
Balance servo calibration failed
6057
Quantity adsorbed calibration failed.
Cause:
Cancel
was clicked while an analysis or a Po measurement was in
progress.
Action:
Restart the operation.
Cause:
An error in the valve control electronics has occurred.
Action:
Contact a Micromeritics service representative if this error message
continues.
Cause:
A power failure occurred during an automatic operation. The
operation was cancelled.
Action:
Restart the operation.
Cause A:
Calibration was out of the tolerance range.
Action A:
Verify the preparatory steps for the calibration were performed
properly. For example, the sample tube may not have been empty.
If all preparatory steps were followed and no other message was
displayed, contact a Micromeritics service representative.
Cause B:
One of the steps in the calibration failed. An additional error message
is displayed.
Action B:
Refer to the Cause/Action for the error message that displayed with
this one.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...