16. Plug the pump power cord into the power source. Allow the pump to run a few hours (overnight if
possible) to eliminate air and moisture from the fresh oil and to produce efficient vacuum
operations.
Related Topics
See "Inspect and Change Vacuum Pump Oil" on page 6 - 3
See "Replace the Vacuum Pump Exhaust Filter" below
R
EPLACE THE
V
ACUUM
P
UMP
E
XHAUST
F
ILTER
The equipment shown in this section may differ slightly from yours.
However, the process is the same unless otherwise noted.
The gases used by the analyzer are exhausted by the oil-filled vacuum pump. The filter minimizes the
release of oil vapor and should be replaced when it becomes so saturated with oil that it is ineffective.
Exhaust filters are used to minimize the release of oil vapors. The gases
are diluted substantially upon release from the vacuum pump; however,
it may be necessary in some locations to provide a fume hood for pro-
tection from hazardous gases and vapors released into the work area.
1. Turn off the power to the pump and disconnect the pump from the power source.
2. Loosen the clamp wing nut at the vacuum pump exhaust port. Swing the clamp away from the
exhaust port and remove it.
Replace the Vacuum Pump Exhaust Filter
Smart VacPrep Operator Manual
6 - 11
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...