Gemini VII
Appendix C
239-42828-01 - Dec 2012
C-1
C. CALCULATIONS
This appendix contains the calculations used in the analysis program.
Saturation Pressure
Saturation pressure (Po) is selected on the Po and Temperature Options dialog. It may be entered or
measured in the Po tube or on port 3.
• If entered, Po = user-entered value
• If measured, Po = equilibrated pressure reading after saturating Po tube or the sample tube
Relative Pressure Calculations
Po
I
Po
1
Po
2
Po
1
–
t
I
to
1
–
to
2
to
1
–
----------------------------------------------------------
+
=
If Po was measured at intervals in the Po tube and several data points were collected in the interval:
• For points taken before the previous Po measurement,
• For points taken after the previous Po, use the previous Po:
Po
I
= Po
1
If Po was entered, use the entered Po:
Po
I
= Po
E
P
rel
1
P
1
Po
I
--------
=
Calculate relative pressure for the I
th
data point:
where
Po
I
=
saturation pressure for the I
th
data point (mmHg)
Po
1
=
most previous measured saturation pressure before I
th
data point (mmHg)
Po
2
=
first measured saturation pressure after the I
th
data point (mmHg)
Po
C
=
calculated Po
Po
E
=
entered saturation pressure (mmHg)
t
I
=
time when the I
th
data point was taken (minutes)
to
1
=
time when Po
1
was measured (minutes)
to
2
=
time when Po
2
was measured (minutes)
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...