Warning: Changing the calibration information will affect the performance of
the instrument. Only qualified service personnel should do this. Do you wish to
proceed?
2516-
Warning: Keeping a backup copy of the calibration data is recommended by
Micromeritics. Would you like to do so now?
2517-
Canceling this dialog will reset the calibration state to what it was when this
dialog was first opened. Are you sure you want to cancel?
2520-
No data points available for reporting.
2522-
Invalid controller application file.
2532-
The instrument contains a different software version. Do you want to reset it?
Cause:
The process of performing a calibration operation was started.
Action:
Calibration operations should only be done by or under the direction
of qualified service personnel.
Cause:
A calibration operation was performed and a backup copy is
recommended.
Action:
Perform a calibration
Save
operation.
Cause:
The calibration has not been accepted.
Action:
If the calibration operation was successful, press
Accept
.
Cause:
The selected sample file does not have collected data and cannot be
used for reporting.
Action:
Select a different sample file.
Cause:
The application’s control file has been corrupted or deleted.
Action:
Reinstall the analysis program.
Cause:
The application has discovered a different version of software
operating in the analyzer.
Action:
If there are no analyzers other than the Gemini connected to the
computer, select
Yes
and allow the updated software to load.
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...