Gemini VII
Appendix B
239-42828-01 - Nov 2013
B-15
4042-
0.0 cannot be a starting or ending pressure for a geometric progression from
low pressure.
4043- 1.0 cannot be a starting or ending pressure for a geometric progression toward
saturation.
4044-
Points in the Langmuir report pressure table lie outside the collected data.
4045-
Points in the report pressure table lie outside the collected data.
4046- [file name] could not be opened for reading.
4047-
Warning: An error occurred while reading [file name].
Cause:
An attempt was made to generate a pressure table from a
geometrically progressing range.
Action:
Change the 0.0 entered value.
Cause:
An attempt was made to generate a pressure table from a
geometrically progressing range.
Action:
Change the 1.0 entered value.
Cause:
Calculation assignments are not being used and more than one of the
report pressure table points is above the range of the collected data
and more than one is below.
Action:
Change the report pressure table to be more consistent with the
collected data.
Cause:
Calculation assignments are not being used and more than one of the
report pressure table points is above the range of the collected data
and more than one is below.
Action:
Change the report pressure table to be more consistent with the
collected data.
Cause:
A thickness curve file could not be opened.
Action:
If the problem persists, restart the computer and optionally perform a
media integrity check (using ScanDisk).
Cause:
An error happened during a read operation of a thickness curve file.
Action:
If the problem persists, restart the computer and optionally perform a
media integrity check (using ScanDisk).
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...