1 A
BOUT THE
S
MART
V
AC
P
REP
The Smart VacPrep is an advanced six-port degassing system that utilizes vacuum to degas samples by
heating and evacuation. Each of the ports may be operated independently of one another. Samples may
be added or removed from degas ports without disturbing the treatment of other samples undergoing
preparation. When starting a new sample, ports already degassing samples are automatically suspended
unless they are in the slow evacuation stage. When unrestricted evacuation is attained by the new
samples, the previously suspended samples will automatically resume their evacuation. Degassing will
automatically terminate when each sample has completed all programmed steps.
The Smart VacPrep remembers the previously selected set of degas conditions and allows them to be
reused. However, the option of programming each port with a different set of parameters is available. A
record of the sample preparation time and temperature can be recorded as part of the sample data file.
The degas program may be started or terminated by using the analyzer software on the computer
attached to the analyzer or with push buttons on the front panel of the Smart VacPrep. The push button
control on each port provides the ability to attach the sample tube with heating mantle, and immediately
start degassing a sample.
There are two methods of starting the degas process from the analyzer software:
l
Start Degas
menu option. Accessed from the
Smart VacPrep > Unit [n] > Start Degas
menu selec-
tion. Provides the option to use degas conditions from selected sample files or a degas conditions files
for each Smart VacPrep port. When using the
Start Degas
menu option, the degas process is started
using the
Start
button on the
Start Degas
window. This option is more convenient for samples that
have specific degas requirements that should be associated with the sample file when it is created.
1 About the Smart VacPrep
Smart VacPrep Operator Manual
1 - 1
Summary of Contents for Gemini VII 2390a
Page 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Page 4: ......
Page 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Page 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Page 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Page 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Page 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Page 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Page 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Page 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Page 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Page 264: ......
Page 266: ......
Page 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Page 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Page 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Page 380: ...Contact Us vi Smart VacPrep Operator Manual...
Page 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...