PROGRAMMABLE DIGITAL I/O CHANNELS
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL A-13
CU A002 15A
MANTA TEST SYSTEMS
Example 2
The following command sequence sets the phase comparison output to 0 in prefault, to 70° 45ms after entry
into the fault state, and to “off” (zero) in the postfault state.
FIA30
<- set fault incidence angle to 30°
DIO,PCO1
<- enable phase comparison output feature
DIO,PCP0,0
<- prefault phase comparison output disabled
DIO,OUT0,0,H8000
<- force prefault value to zero
DIO,PCP1,1,70
<- fault phase at 70°
DIO,PCD45
<- change to fault phase 45ms after entry into fault state
DIO,PCP2,0
<- postfault phase comparison output disabled
DIO,OUT2,0,H8000
<- force postfault value to zero
A.7 COMMAND SUMMARY CHART
* Indicates multiple definitions possible.
Digital I/O Command Summary
Start trigger channel
DIO,STR#
Stop trigger channel
DIO,STP,#
PREFAULT
FAULT
POSTFAULT
Outputs [value, mask]
DIO,OUT0,#,#
DIO,OUT1,#,#
DIO,OUT2,#,#
Delayed Outputs [time, value, mask]
*DIO,ODY#,#,#
.
.
.
Conditional Outputs
[input value, input mask, delay, output value,
output mask]
*DIO,OCD#,#,#,#,#
.
.
.
Phase Comparison Phase [0/1, deg]
DIO,PCP0,#,#
DIO,PCP1,#,#
DIO,PCP2,#,#
Phase Comparison Delay [ms]
DIO,PCD#
Phase Comparison Output
Misc. Output Control
Enable/Disable [1/0]
DIO,PCO#
Output settings clear
DIO,OCL
Width [us]
DIO,PCW#
Read outputs
DIO,ORD
Other Input Commands
General Commands
Sequence of events report summary
DIO,SEQ
Value & mask print format
DIO,FMT#
Sequence of events report
DIO,SER
Status
DIO,STA
Read Inputs
DIO,INP
Postfault inputs
DIO,POI