DETAILED OPERATION - Section 4
4-50
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
CONTINUOUS-ADJUST
OFF 24v 48V 125V 250
Continuous adjustment of DC voltage output:
The DC voltage output can be varied continuously from 24V to 300V by selecting SETTINGS DC-VOLTS
CONTINOUS-ADJUST in the menu. The [MODE/MENU DISPLAY] will appear as shown above.
Turn the [MODIFY] knob to adjust the voltage. Press [MODE/MENU] or [SELECT] when done.
DC VOLTAGE OUTPUT:
48.0 VOLTs
For units with the additional output enable button and LED indicator, when the output is disabled, the value
in the [MODE/MENU DISPLAY] is the setting which will be applied when the output is enabled. You’ll
note that the measured value is basically zero in the [VALUE DISPLAY] if the output is disabled.
Note: The DC voltage output isn’t changed when the set defaults action is executed using the SETTINGS
DEFAULT menu selection, or the RS-232 DEF command. This prevents resetting of solid state
relays during testing, which are supplied from the DC voltage output.
See Section 6.2.15 for instructions on controlling the DC voltage output via the RS-232 port.
Setting the Default DC Voltage:
The default DC voltage output on power up can be set by selecting OTHER DEFAULT DC-VOLTAGE in
the menu.
Turn the [MODIFY] knob to adjust the voltage. Press [MODE/MENU] or [SELECT] when done.
The default setting on power-up also can be set via the RS-232 interface using the DDF# command. See
Section 6.2.15.