SERVICING - Section 8
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 8-13
CU A002 15A
MANTA TEST SYSTEMS
2.
Discharge any static buildup by touching all tools and hands to the case. Wear a proper grounding
wristrap.
3.
Remove the top cover, and unplug the card in the first plug in-slot from the front. Locate the lithium
battery in location B1. Remove the battery.
4.
Replace with an Eveready CR2032 or equivalent battery. Note that the “+” side of the battery must face
up, and the “-” side must face down.
5.
Plug the card firmly back into the first plug-in slot, and restore any disconnected connectors.
6.
Check to see that all plug-in cards are fully seated. They should all be at the same height. Replace the
top cover.
7.
Turn on the MTS-1710. With a computer connected to the COM1 port of the instrument, send the RTR
command using a communications program, such as PowerTerm or PROCOMM. Then set the clock
using the MDY#,#,# and HMS#,#,# commands. These commands are described in Section 6.2.12.
8.
Check the clock setting using the TIM command, or via the OTHER CLOCK selection in the front panel
menu.
8.10 LED/LAMP TEST FUNCTION
All the LEDs and lamps on the front panel can be tested using the OTHER SERV LED-TEST. This turns
on all LEDs and lamps, except for the [TONE] pushbutton LED.
To check the [TONE] pushbutton LED, short the [EXTERNAL STOP] trigger contact sense input
terminals, and this button should light.