MTS-1720 TWO CHANNEL CURRENT SOURCE - Section 9
9-24 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
Example 4 The following example command sequence programs the two-phase-to-ground fault, as shown
in Figure 9.11.
FMD0
<- select
Φ
-N fault type to set nominal amplitudes
IMD5
<- I3-WYE current mode
DYN
<- dynamic operation mode
AVA70
<- set prefault
Φ
-N voltage amplitudes to 70V
AVB70
(usually these already have been set as default values)
AVC70
FMD13
<- 2
Φ
-N B-C fault mode
AIA1
<- set prefault
Φ
-N currents to 1A
AIB1
AIC1
PRF1
<- prefault on
POF0
<- postfault off
<- delay two seconds here for phase reading before presetting phase
PRP
PHS10
<- prefault phase angle 10
°
PFI70
<- fault phase angle 70
°
IFI6
<- initial fault current 6A (10.4A phase-to-phase)
VPR70
<- prefault 2
Φ
-N voltage at 70V (120V phase-to-phase)
VFI70
<- fault voltage 70V (120 phase-to-phase)
IPR1
<- 2
Φ
-N prefault
Φ
-N currents at 1A
STR
<- initiate fault
Example 5 The following example command sequence programs a three-phase fault involving a current
and voltage ramp. Postfault is set on, with an auto-reclose delay of one second.
FMD0
<- select
Φ
-N fault type to set nominal amplitudes
IMD5
<- I3-WYE current mode
DYN
<- dynamic operation mode
AVA70
<- set prefault
Φ
-N voltage amplitudes to 70V
AVB70
(usually these already have been set as default values)
AVC70
FMD6
<- 3
Φ
A-B fault mode
IPR1
<- 3
Φ
prefault current 1A (1.73A phase-to-phase)
PRF1
<- prefault on
POF1
<- postfault on
<- delay two seconds here for phase reading before presetting phase
PRP
PHS10
<- prefault phase angle 10
°
PFI70
<- fault phase angle 70
°
VPR100
<- prefault voltage 120V (100% of 120V nominal)
VFI80.0
<- initial fault voltage 96V (80.0% of 120V nominal)
VFF28.8
<- final fault voltage 34.6V (28.8% of 120V nominal)
VRR25
<- voltage ramp rate at 25%/sec
VPO100
<- postfault voltage at 120V (100% of 120V nominal)
IFI6
<- initial fault current 6A (10.4A phase-to-phase)
IFF15
<- final fault current 15A (21.2A phase-to-phase)
IRR4
<- ramp currents at 4A/sec