ADVANCED OPERATION - Section 5
5-10 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
5.2.8 Programming Durations
Durations from 0-99.999 seconds (6000 cycles at 60 Hz) can be programmed. They can be programmed in
either seconds or cycles (based upon the present output frequency).
While adjusting a duration setting in the [MODE/MENU] display, pressing the [TIME] button toggles
between display in seconds and cycles.
Example duration setting in seconds:
VOLTAGE DURATION
2.972 SEC
Pressing [TIME] changes the display to cycles (shown here assuming 60 Hz output):
VOLTAGE DURATION:
178.2 CYCLES
5.2.9 Example Applications of Ramp and Fault Duration Features
Example 1
The following settings map shows an example simulation of a phase-to-phase fault.
VAB is ramped from the nominal 120V to 10V, and the current is stepped from 1A to 4A, then ramped to
12A. The phase angle VAB to IAB is stepped from 12
°
in prefault, to 75
°
during the fault.
MTS-1710 Settings Map Example
Fault Type
Φ
-
Φ
Operation mode
DYNAMI
C
Freq. reference mode
LINE
Fault Phase
A-B
Current mode
I3
Harmonic
1
Parameter
PREFAULT
FAULT
POSTFAULT
ON
Initial
Ramp Rate
Duration
Final
OFF
VAN [V]
69.3
-
-
-
-
-
VBN [V]
69.3
-
-
-
-
-
VCN [V]
69.3
-
-
-
-
-
VAB [V]
120.0
120.0
20%/sec
off
10.0
N/A
I1 or I3[A]
1.0
4.0
3A/sec
off
12.0
N/A
Phase [deg]
12.0
75.0
off
off
N/A
N/A
Freq. [Hz]
N/A
N/A
off
off
N/A
N/A