SERVICING - Section 8
8-8 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
4. Set the dynamic measurement mode to PEAK (see section 5.1.1), and the operation mode to
DYNAMIC. Press [FAULT].
Repeat steps 2 to 3 to adjust the peak measurement calibration factors.
IMPORTANT: This must be done with the dynamic measurement mode set to PEAK, and in the
FAULT state (FAULT button flashing).
To avoid overheating of the output leads while performing this calibration, you can use STATIC operation
mode. Just remember to press and hold the [FAULT] button when actually adjusting the calibration factor.
5. When done, the CAL1 command must be sent via the RS-232C interface to enable saving of the
calibration factors.
Next, select OTHER SERV CALIBRATION SAVE to save the calibration factors in non-volatile memory
so that these new calibration factors will be remembered when the MTS-1710 is powered up next time.
8.5.4 DC Current Measurement Calibration
1.
Set dynamic measurement mode to RMS (see section 5.1.1). Select I4 current mode, and connect an
external reference DCammeter to the I4-DC OUTPUT terminals.
2.
Press [CURRENT], adjust the current down to 0, and then up to 2.0A (as read on an external reference
meter).
Select OTHER SERV CALIBRATION MEASUREMENT I4 in the menu. Turn the [MODIFY] knob
to adjust the calibration factor until the reading is the same as read on a reference meter (the output
current doesn’t change when this is done).
Press [SELECT].
3. When done, the CAL1 command must be sent via the RS-232C interface to enable saving of the
calibration factors.
Next, select OTHER SERV CALIBRATION SAVE to save the calibration factors in non-volatile
memory so that these new calibration factors will be remembered when the MTS-1710 is powered up
next time.
Note: If you’re performing all calibrations, only one SAVE operation at the very end is required.
8.5.5 AC Current I3 Measurement Calibration (
Φ−
N
Values)
(This applies only to systems with an MTS-1720).
1.
Remove all loads from the current output terminals.
2.
Set dynamic measurement mode to RMS (see section 5.1.1). Select I3-WYE current mode, and then
connect an external reference ammeter to the I3 OUTPUT B-N terminals.
3.
Press [CURRENT], adjust the current down to 0, and then slowly up to 6.0A (as read on an external
reference meter).