DETAILED OPERATION - Section 4
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 4-27
CU A002 15A
MANTA TEST SYSTEMS
4.6.10.4.2
Three phase connection
This connection is for three-phase testing with a single phase current (up to 60Amax, 800VA max, 44Vrms
max), using a three-phase, four-wire connection.
FIGURE 4.19 PARALLELING FOR A THREE-PHASE LOAD
For current paralleling:
1.
Connect the multi-system synchronization cable at the rear, as shown in Figure 4.25.
2.
Select I3 current mode, and select the same fault type and phase on both MTS-1710s.
3.
Connect the MTS-1710 current outputs, as shown in the diagram above.
4.
To synchronize the fault initiation/termination, connect the start/stop trigger inputs, as shown.
5.
Set the current amplitude and phase angle on each MTS-1710/1720 system individually (normally set
to the same on both). The current through each phase of the load will be the vectorial sum of the
paralleled currents.
TRIP
THREE-PHASE RELAY
I3
Multi-system
sync connection
(rear)
Note: Select I3 current mode on both MTS-1710 systems
IN
IC
IB
IA