RS-232C INTERFACE - Section 6
6-28 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
6.2.15 DC Voltage Control
This section only applies to units with the DC Voltage output option.
The following commands are associated with control of the DC voltage output:
Command Description
VDC#
Volts DC
•
Set DC voltage output
•
Valid values are 0, and 24 to 300 Volts
•
A value of zero disables the DC voltage output
•
Enables DC output if the value is non-zero
PVD
Print Volts DC
•
Prints the present value of DC voltage (in volts)
DVD Display
Volts
DC
•
Selects DC voltage on the [VALUE DISPLAY]
DDF#
DC Voltage Default
•
Sets the default DC voltage on power-up
•
Valid values are 0, and 24 - 300 (Volts) (250V maximum may apply on older
MTS-1710 systems)
•
This value will take effect the next time the MTS-1710 is turned on
6.2.16 COM2 Interface Commands
The following commands sent to the MTS-1710 COM1 port are used to control or send/receive data from
the COM2 port:
Command
Description
C2B#
COM2 Baud Rate
•
Sets the baud rate for COM2
•
Valid values are 0-12. The default value is 11 (9600 baud).
0 - 50 baud
7 - 1050 baud
1 - 110 baud
8 - 2400 baud
2 - 134.5 baud
9 - 4800 baud
3 - 200 baud
10 - 7200 baud
4 - 300 baud
11 - 9600 baud
5 - 600 baud
12 - 38400 baud
6 - 1200 baud
•
The change in baud rate is effective immediately following execution of the
C2B command.
•
The new setting is retained in non-volatile memory, and takes effect until
changed again.